Test and Prompt List
5-2 Pulse oximeter Service Manual
5.2 Prompt List
Table 5-1 Prompt Information on LCD
Indications Cause Solution
Low battery
symbol
The battery voltage is
below the threshold value
Replace the battery
Memory full
symbol
Available data memory
locations <10
The existing data will be overwritten.
Export the data in time.
Memory full
symbol blinks
The memory is full
The existing data has been overwritten.
Export the data in time.
ID full symbol ID > 95
ID will be overwritten. Export the data in
time.
ID full symbol
blinks
ID has been overwritten
ID has been overwritten. Export the data
in time.
Standby symbol Standby mode None
Communication
symbol
Communication mode None
DELETE ALL? Delete button pushed None
ALL DELETED
Delete button pushed
again after “DELETE
ALL?”
None
Table 5-2 Error indications
Error Message Cause
Solution
Initiate Error
Failed self-test
Shut down the device (if can’t, remove the
batteries) and contact Mindray DS for service.
Please Release
the Button
Button error
Check for jammed button. If problem remains,
contact Mindray DS for service.
Pulse Not Found
Searching…
Pulse not
found
Check the patient and alert the doctor.