9-12
9.1.5 Event Review
9.1.5.1 Overview
Clicking the “Event Review” tab will enter the following tab sheet, through which you can
view all event parameters and waveforms of a patient.
1 Event list 2 Event parameter area 3 Event waveform area
Figure 9-4 Event Review Tab Sheet
1. Event list: displays event status (locked or not), time, message, priority, and description.
2. Event parameter area: displays all parameter values for each event time. The parameters
from integrated devices are indicated with “+”before the parameter name, and the
relevant text is shown in white. The CMS prompts if an alarm occurs to a specific
parameter and indicates the alarm priority by using different background colors.
Black: no alarm generated;
Red: high level alarm generated;
Yellow: medium or low level alarm generated.
3. Event waveform area: displays relevant parameters’ waveforms within ±16 seconds
around the event time. You can use caliper to measure the time difference between any
two points on waveform. Please refer to 9.1.3.3 Normal Waveform Review- Waveform
Area for how to use caliper.
Select an event from the event list, and you will view its corresponding parameter values and
waveforms in the event parameter area and event waveform area.
NOTE
z For the event from integrated devices, there is no corresponding waveform in
the event waveform area.
1
3
2