6-6
Night Mode: Open the “Multibed” menu, select and set current bed or all beds to Night
Mode, and choose whether to stop NIBP measuring.
ST Display Setup: Access the “ST Display Setup” of the telemetry system.
Start/Stop NIBP Measurement: Used to start or stop NIBP measurement.
Move zone to: Selecting it to switch over the current patient window to any destination
window.
NOTE
z One type of monitor can support only one of the following functions: Monitor
Silence, Audio Paused, and Silence (Please refer to the corresponding operation
manual for more information). Therefore, these features’ corresponding
controls and buttons do not exist at the same time.
z Any of the operations of Monitor Silence, Audio Paused and Silence will set the
corresponding patient windows on the CMS to silence status.
6.2.3.3 Alarm Icons
No icon: Indicates that the system sound of the monitor is normal.
: Indicates that the alarm of the monitor is paused.
: Indicates that the alarm of the monitor is turned off.
: Indicates that the system sound of the monitor is silenced.
: Indicates that the alarm sound of the monitor is switched off.
: Indicates that the alarm sound of the monitor is paused or the monitor is in alarm
silence status.
is the icon of defibrillator, which indicates that a defibrillator is currently
connected to the bed.
NOTE
z When an alarm icon appears, the defibrillator icon will not be displayed.