12-2
Set Unit
You can set unit for each parameter of the following modules as required:
Module Parameter Options Defaults
IBP IBP mmHg, kPa, cmH2O mmHg
CO2 CO2 mmHg, kPa mmHg
TEMP T1, T2, TD, TB
,
Vascular Resistance SVR/PVR DS/cm
5
, kPa-s/l DS/cm
5
NIBP NIBP mmHg, kPa mmHg
CO2 mmHg, kPa, % mmHg
GAS
O
2
mmHg, kPa, % mmHg
ECG ST mV, mm mV
Oxygenation Hb g/dl, g/l, mmol/l g/dl
Telemetry Alarm Delay Setup
You can set telemetry alarm delay time (6 sec by default) and ST alarm delay time (30 sec by
default). If the alarm-triggered condition disappears within the delay time, the CMS will not
give the alarm. The alarms include: HR too high, HR too low, SpO2 too high, SpO2 too low,
PR too high, PR too low, ST too high and ST too low.
Draw Waveform Type
There are two options:
COLOR: Selecting this option allows you to see smoothed waveforms.
MONO: Selecting this option allows you to see unprocessed waveforms.
Display Grid
You can set whether to display grid in the waveform area.
Enter the Admin Setup tab sheet
By clicking the “Admin Setup button and inputting the required password, you can enter the
Admin Setup window. For details, refer to part 12.2 Admin Setup.
Enter the Factory Setup tab sheet
By clicking the “Factory Setup” and input the required password, you can enter the Factory
Setup window. As the “Factory Setup” is intended for use only by the
manufacturer-designated personnel, no details will be given to it in this manual.