5-10
NOTE
z Transferring patient and combining with destination bed patient data may cause
some patient data loss if the system time is inconsistent between the source and
destination CMS.
z Ensure that the same patient is monitored by both the source and destination beds
when transferring patient and combining with destination bed patient data.
Otherwise, select the function of discharging destination bed patient with data
saved.
z When a patient is transferred from one bed to another, the CMS will interrupt
bedside monitoring temporarily. However, whether to continue monitoring the
patient depends on the doctor's decision.
z When a patient is transferred within one CMS, all the monitored data will be
transferred along; if the patient transfer is among different CMSs, only the
full-disclosure waveforms and other data of the latest 3 days are transferred along.
z If there is version difference between two CMSs, patient transfer may not be
enabled.
z When a defibrillator is connected to the CMS, patient transfer is not enabled.
5.8.2 Transferring a Patient on Monitor
When a patient is transferred between monitors by a transferring device, the monitor will
inform the CMS of the transference. The CMS will automatically operate as follows:
Discharge the patient from source bed;
Combine the patient data at source bed with the data at destination bed.