12-16
NOTE
z The "Obtain Patient Information" button is available on the patient
management screen only when“ADT Query”is set to “Enable”.
Date Export Setup
Input multicast IP address and port.
Export Waveform Data: When configured to “ON”, the button “Export to EMR”
will be displayed on the Waveform Review and Alarm Review sheets and the
waveform or event can be exporte
d to other systems. When configured to “Off”,
the waveform or event can not be exporte
d to other systems.
Multi-CMS Communication Settings
Multi-CMS Communication Settings menu is used to add or delete the IP address of other
CMS.You can view other beds on that CMS only when the corresponding IP address of a
CMS is added.
1. Click the “Multi-CMS Communication Settings” button and open the “Communication
Settings” menu.
2. Click “Add”, and then input the IP address of a CMS on which you want to proceed
viewing other beds in the opened dialog box.
3. Click “OK”.
NOTE
z As for the CMSs or monitors which only have broadcast mode to implement
remote CMS, it is required to tick the “Broadcast send mode” checkbox.
Kommentare zu diesen Handbüchern